been reports of sensors causing patient burns when
operating in an MRI environment.
Do not use the Monitor in the presence of flammable
anesthetics.
To help prevent unintended current return paths with
the use of high frequency (HF) surgical equipment,
ensure that the HF surgical neutral electrode is
properly connected.
To avoid personal injury, do not perform any servicing
unless qualified to do so.
WARNING: These Monitors should not be used on
patients who are connected to cardiopulmonary bypass
machines.
If powering the Monitor from an external power
adapter or converter, use only power adapters and
converters approved by GE Medical Systems
Information Technologies.
The Monitor does not include any user-replaceable
fuses. Refer servicing to qualified service personnel.
To reduce the risk of electric shock, do not remove the
cover or the back. Refer servicing to a qualified service
person.
If the accuracy of any determination reading is
questionable, first check the patient’s vital signs by
alternate means and then check the PRO Monitor for
proper functioning.
Cautions
Do not use replacement batteries other than the type
supplied with the Monitor. Replacement batteries are
available from GE Medical Systems Information
Technologies. See Appendix D.
The PRO Monitor is designed to conform to
Electromagnetic Compatibility (EMC) standard IEC
601-1-2, 1993 and will operate accurately in
conjunction with other medical equipment which also
meets this requirement. To avoid interference problems
affecting the Monitor, do not use the Monitor in the