If there is excessive light, cover the sensor with opaque
material.
PROBLEM: A sensor error indicating a bad sensor has
occurred.
CAUSE:
The sensor or cable may be defective, or the cabling
may be improperly connected.
SOLUTION:
Check the patient.
If possible, keep the patient still; check whether the
sensor/cable is applied securely and properly, and
replace it if necessary.
Disconnect and reconnect the sensor.
NELLCOR
®
OxiMAX SpO
2
Description
The SpO
2
parameter in the PRO Monitor is available in two
different leading technologies: NELLCOR
®
OxiMAX and
MASIMO SET
®
. Please refer to the front of your Monitor to
see which SpO
2
technology your monitor contains. The
SpO
2
technology logo will be on the front fascia of the
Monitor. This section refers to NELLCOR SpO
2
technology.
Functional oxygen saturation (SpO
2
) of arterial blood is
noninvasively and continuously monitored in the PRO
Monitor using pulse oximetry technology from NELLCOR
®
.
Functional SpO
2
is the ratio of oxygenated hemoglobin to
hemoglobin that is capable of transporting oxygen.
Pulse rate derived from SpO
2
appears in the Pulse BPM
display (22), and the SpO
2
pulse indicator (20) flashes
synchronization with the real-time pulse rate measurements
*NELLCOR
®
is a trademark of Nellcor Puritan Bennett