Getting Started
in a medical/hospital grade attachment plug, provided
with the following cord tag: “Hospital Grade Plug."
Grounding integrity can only be maintained when
equipment is connected to an equivalent receptacle
marked "Hospital Grade."
• Where the integrity of the external earth conductor in
the installation or its arrangement is in doubt, the
Monitor must be operated from its internal battery.
General Caution
Do not touch either the pin of the DC input connector
(3) or the terminals within the battery compartment (1)
and the patient at the same time.