Alarms Button
Selection of this button displays the Alarms menu. This menu
is used to adjust the violation limits for BP, Pulse Rate, and
SpO
2
. The values and ranges for these parameters are not
stored when the Monitor is turned off. The user may edit the
limits, but they are restored to the default values each time
the Monitor is switched on. To permanently change the
alarm limits, refer to “Alarms” under “Service Button” in the
“Using the Menu System” section.
Volume
Selection of this button displays the alarm volume submenu.
The volume range is from 1 to 10, with 10 being the loudest.
The alarm volume is stored when the Monitor is turned off
and restored to the user’s preference each time the Monitor
is switched on. Selection of the Check button allows the
current volume setting to be heard. Selection of the Main
button returns the user to the Main menu.
Alarm Limits Table
Parameter Range Default
Systolic High 35 - 245 180
Systolic Low 30 - 240 30
Diastolic High 15 - 195 130
Diastolic Low 10 - 190 15
MAP High 20 - 215 140
MAP Low 15 - 210 50
BPM High
(Masimo)
35 - 235 160
BPM Low (Masimo) 30 - 230 40
BPM High (Nellcor) 35 - 250 150
BPM Low (Nellcor) 30 - 245 50
SpO2 High 21 - 100 Off
SpO2 Low 20 - 99 90